Cutera XEO Error Codes

cutera-xeo

The following list below is a list of Cutera’s error codes for the Cutera XEO machine.  We have the expertise and knowledge to assist you with repairing any issues that cause error codes on your Cutera systems and handpieces.  We also repair other Cutera machines.  If you need help with error codes for other Cutera machines (Cutera Solera, Excel-V, Cool Glide, Vera Lite, and more) please contact us for more information.

Cutera XEO Error Codes

FAULT
ERROR CODE
WHAT IT MEANS
HVPS

HVPS
201

202
BOOST FAULT, POWER SUPPLY

B+ and B+/2 are more than 20 V apart
HVPS
211
The BLEED_FAULT_Q is set
HVPS
213
IGBT_FAULT_Q is set
safety
214
CROWBAR_FB_Q is set
HVPS
215
DC_FAULT_Q is set
HVPS
temperature
217
221
UNEXPECTED POWER
SETP2 is active, water temp >65 degrees C
temperature
224
TipTemp is greater than 20 (Laser) or 30 (Flash) degrees C in READY or EXPOSURE
spotsize
225
Spotsize changed during exposure
Park sensors
226
Selected HP was parked during exposure
Temp Sensor
229
Wires to temp sensor are connected opposite.
Energy fluence
230
Main detector accumulated energy is less than 50% of the requested energy
Energy
231
Main detector accumulated energy is more than 150% of the requested energy
Energy
232
Error lamp over energy
Energy
235
Flash HP efficiency is low replace HP soon
Energy
237
Firepulse fault not all micropulses were delivered or the footsw was released during Titan exp
reference voltage
240
Reference voltage is less than 80% or greater than 120% of expected (10V)
reference voltage
241
Reference voltage is less than 80% or greater than 120% of expected (Vcc)
reference voltage
242
Reference voltage is less than 80% or greater than 120% of expected (5 V)
lamp
243
After the laser lamp is lit, the lamp voltage goes out of range
reference voltage
244
Reference voltage is less than 80% or greater than 120% of expected (15 V)
lamp
246
Flash HP Lamp failed to start
lamp
248
The flash lamp current sensor reports no current
lamp
249
Laser Lamp failed to start
trigger switch
250
Trigger switch (hand button or footswitch) is malfunctioning
shutter
251
Shutter won’t open
shutter
252
Shutter won’t close
communication
270
No response from DISPLAY board in 400 ms
Shots expired
288
Shots expired, must reset chip.
Footswitch Use
289
The footswitch was released during a Titan pulse
Capacity
293
1000 shots left in HP
Capacity
294
2000 shots left in HP
lamp
301
HVPS wont boost check fuse
HVPS
303
Titan HPP mode fault
lamp
310
HVPS fault
HVPS
312
BOOST_OVP_Q is set
HVPS
316
I_LMP_FAULT_Q is set
HVPS
317
INADVERT_EXP_Q is set
HVPS
318
MODE_FAULT_Q is set
shutter
319
SHUTTER_FAULT_Q is set
water flow
322
Pressure switch is not closing when pump is engaged.
temperature
323
S/W cannot read temperature of the Laser handpiece.
temperature
327
S/W cannot read temperature of the Flash handpiece.
Low efficiency
332
Head power is low.
Detector Mismatch
333
Main detector signal missing – check connections – w/E367 check for detector leakage
Detector Mismatch
energy
334

335
Laser Main and safety energy detectors do not agree.

Limelight or prowave efficiency is low.
ADC fault
336
Analog to Digital converter failed
Firepulse fault
337
The exposure was aborted before all micropulses were delivered
safety
338
Detector integrator gain hardware fault check switches in SW2 block
detector mismatch
339
Flash HP Main and safety energy detectors do not agree.
lamp
345
The flash lamp went out after a pulse lamp might be broken
lamp
347
The flash lamp current sensor reports current when the laser lamp is on
NVRAM
360
NVRAM checksum doesnt match or NVRAM version number doesn’t match
calibration
361
During the test fire, the system did not meet minimum efficiency limit.
calibration
362
Any error occurred during POST light reg controlled test fire
PROM
363
PROM checksum doesn’t match
safety
364
POST fired the crowbar and the CROWBAR_FB_Q is not set after 20 ms
safety
365
TipPIC failed its POST most likely leakage current in the tip thermistor
NVRAM
366
The service mode calibration was not completed
safety
367
POST safety circuit test failed
NVRAM
368
The Touch Screen calibration was not completed
calibration
369
POST test fire did not finish in 20 ms
communication
371
Handpiece communication protocol failure
Calibration
372
Model not programmed to either Xeo or Genesis Plus
communication
380
Communication timeout talking to the Flash HP
calibration
381
Data in the Flash HP is unrecognized
calibration
382
Flash HP type is unknown
communication
383
Flash HP shot count info has a corrupted digital signature resave shot data using ServiceTool
communication
384
Communication failure talking to the Flash HP
communication
385
Flash HP iButton data CRC fails
Calibration
386
Xeo HP is not authorized on this machine check enabled HPs with ModelMaker
Calibration
387
Xeo HP is not calibrated
Software
390
Assertion failure software logic problem
Relay
391
Xeo HP relay fault
Software
392
Software version mis-match check the Display flash / U5 versions for correctness

Additional Error Codes:

400    NNY    Reservoir vacuum fault
401    YNY    Water pump off
501    YNY    Power Supply state mismatch
502    NNY    Power Supply rejected command
550    NYY    Scanner X axis fault
551    NYY    Scanner Y axis fault
552    NYY    Scanner shutter fault
553    YNY    Scanner ADC test fault
554    YNY    Scanner power on self test failure
555    YNY    Scanner home not set
600    NNN    User warned!!!
700    NNY    Energy 20 percent high
701    NNY    Energy 20 percent low
999    YYY    Wrong internal data